Digital pH meter delivers precision while offering simplicity and versatility in a sleek design.
HN/002002-02 - "edge" Dedicated pH/ORP Meter
"edge"digital pH meter delivers precision while offering simplicity and versatility in a sleek design.
Your edge meter will automatically recognize each digital electrode, providing sensor type, calibration data, and serial number.
Advanced pH electrode diagnostics ensure worry-free measurements.
Use your edge pH meter as a portable, benchtop meter, or mount it on the wall to save space.
- The large display and it’s wide 150° viewing angle provide one of the easiest to read LCDs in the industry
- Set up and calibrate your meter in no time with 1 point calibration
- Great for use in business and educational laboratories worldwide
"edge" is incredibly versatile.
It can be used in a variety of ways and in a variety of environments without taking up the space of a traditional benchtop meter.
|
CAL Check "EDGE's" exclusive CAL Check feature analyses the pH electrode response in the pH buffers during the calibration process to alert the user of potential problems such as contaminated buffer and clean electrode. After calibration, indicators for probe condition and response time are displayed on the measurement screen. The probe condition indicator is based on offset and slope characteristics of the pH electrode. |
|
Digital pH Electrodes edge pH electrodes are highly advanced, featuring a built-in microchip that stores calibration information that is automatically retrieved by edge once the electrode is plugged in. With a digital electrode, it is possible to change electrodes without having to recalibrate since the calibration information is stored in the pH electrode and not the pH meter. This allows flexibility for the user that measures pH with more than one electrode. (ORP electrode sold separately). |
|
Two USB Ports edge includes one standard USB for exporting data with a flash drive and a second micro USB port for you to connect to a computer for file export and for charging your edge when the cradle is not available. |
|
Wall Mount Using the wall mount cradle (included), edge can be placed on a wall, leaving zero footprint on the benchtop space. The cradle has a built-in connector to power edge and charge its batteries. edge's zero footprint is designed to save you valuable benchtop space.
|
|
Cradle and Electrode Holder edge is equipped with a benchtop cradle with an adjustable swivel electrode holder to charge and hold the edge securely in place at the optimum viewing angle. |
|
Data Logging edge digital pH meter allows you to store up to 1000 log records of data. Data sets include readings, GLP data, date and time. |
|
Two Operating Modes edge can be used in Extended or Basic Operating Modes. Extended Mode enables all features available while the Basic Mode reduces the amount of features to a very simple operation. For example, in extended mode edge offers 5 point calibration with 2 custom buffers while in basic mode calibration can only be performed to 3 points with 5 preprogrammed buffers.
|
|
Sensor Check When used with electrodes equipped with a matching pin, edge constantly checks the impedance of the pH measuring electrode to notify you in real time in the event of glass breakage. During calibration, Sensor Check checks the state of the junction. The reference junction is also evaluated and reported on the display. |
pH Specifications
pH Range |
-2.000 to 16.000 pH (Standard Mode), -2.00 to 16.00 pH (Basic Mode) |
pH Resolution |
0.001 pH, 0.01 pH |
pH Accuracy (@25ºC/77ºF) |
±0.01 pH, ±0.002 pH |
pH Calibration |
5 points (Standard mode) 1.68, 4.01 (3.00†), 6.86, 7.01, 9.18, 10.01, 12.45, and two custom buffers; 3 points (Basic mode) 4.01; 6.86; 7.01; 9.18; 10.01 |
pH Temperature Compensation |
ATC: -5.0 to 100.0ºC; 23.0 to 212.0°F* |
mV Range |
±1000.0 mV; ±2000.0 mV |
mV Resolution |
0.1 mV |
mV Accuracy |
±0.2 mV (±999.9 mV); ±1 mV (±2000 mV) |
Relative mV Calibration |
Single point calibration |
Temperature specifications
Temperature Range |
-20.0 to 120.0 ºC, -4.0 to 248.0 °F |
Temperature Resolution |
0.1 °C, 0.1 °F |
Temperature Accuracy |
±0.5 °C, ±0.9 °F |
°C/°F |
Yes |
Additional specifications
pH Electrode Diagnostics |
Glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time |
GLP |
Yes |
Logging |
up to 1000*** records organized in: Manual log-on-demand (Max. 200 logs), Manual log-on-stability (Max. 200 logs), Interval logging *** (Max. 600 samples; 100 lots) |
Connectivity |
1 micro USB port for charging and PC connectivity, 1 USB port for storage |
Environment |
0 to 50°C (32 to 122°F), RH max 95% non-condensing |
Battery Type/Life |
Built-in rechargeable battery with up to 8 hours of continuous use |
Power Supply |
5 VDC adapter (included) |
Dimensions |
202 x 140 x 12.7mm (8” x 5.5” x 0.5”) |
Weight |
250 g (8.82 oz) |
Ordering Information |
edge® HN/002002-02 digital pH meter includes benchtop docking station with electrode holder, wall mount cradle, USB cable, HN/011310 glass body refillable pH electrode with temperature sensor, 2 sachets each of pH 4, 7, 10, cleaning solution, 5 VDC power adapter and instruction manual. |
Notes |
* temperature limits will be reduced to actual probe/sensor limits *** For standard mode only (logging) † pH 3.00 is only visible when using specific pH electrodes and will replace pH 4.01 buffer
|