Electronic pipettes ML Series
Based on stepper motor with microprocessor controlled piston movement
Electronic Micropipettes ML Series
This innovative pipette is based on stepper motor with microprocessor controlled piston movement.
These are user friendly and offer effortless pipetting.
Accuracy & precision levels are very high and user independent. This is a complete solution for every pipetting need.
- User friendly software.
- Easy, user friendly calibration.
- Recharging during operation.
- Optimized ergonomy and working comfort.
- Large liquid crystal display (LCD).
- Built-in tip ejector.
- Six models cover the complete pipetting range from 0.2 µl to 1000 µl.
Features
- Model (Easy selection of modes from a single key)
- Standard mode
- Stepper mode
- Dilution mode
- Calibration mode
Variable Working Speeds
- Speed can be selected on a scale of 1-5
Calibration
- User friendly calibration mode facilitates easy calibration
Battery
- 9V NIHM rechargeable batteries are used
- 8 hour of operation with fully charged baatteries
Volume Selection Easy volume selection by 'UP' & 'DOWN' keys.
|
Large Fire Button Single 'FIRE BUTTON' for filling and dispensing.
|
Large Liquid Crystal Display Two Line Alphanumeric LCD.
|
Built-in streamlined tip ejector For taken easy tip ejection and access to narrow naked bottles.
|
Specifications & Ordering Information
Model |
Volume Range |
Inc. µl |
A |
CV |
±% |
±µl |
±% |
±µl |
ME-10 |
0.2-10 µl |
0.05 |
1.2 |
0.12 |
0.8 |
0.08 |
ME-120 |
5-120 µl |
0.5 |
0.8 |
0.96 |
0.3 |
0.36 |
ME-300 |
20-300 µl |
1.0 |
0.8 |
2.4 |
0.3 |
0.9 |
ME-1000 |
50-1000 µl |
5.0 |
0.8 |
8 |
0.3 |
3 |
ME-5000 |
100-5000 µl |
10.0 |
0.8 |
40 |
0.3 |
15 |
Error limits according to the nominal capacity (=maximum volume) indicated on the instrument, obtained with instrument and distilled water at equilibrium with ambient temperature at 20 0C, and with smooth, steady operation. The error limits are well within the limits of DIN EN ISO 8655.2.
(See table-1) A= Accuracy, CV= Coefficient of variation.
Table 1 - Maximum permissible errors as per ISO 8655-2
Nominal Volume |
Maximum Permissible Systematic Error ( ISO 8655-2) |
Maximum Permissible Systematic Error (ME Series Pipettes) |
±% |
±µl |
±% |
±µl |
10 |
1.2 |
0.12 |
0.8 |
0.08 |
100 |
0.8 |
0.8 |
0.3 |
0.3 |
200 |
0.8 |
1.6 |
0.3 |
0.6 |
500 |
0.8 |
4.0 |
0.3 |
1.5 |
1000 |
0.8 |
8.0 |
0.3 |
3.0 |
5000 |
0.8 |
40 |
0.3 |
15 |