Autoclaves IC-T Series Class B
The sterilization solution for larger instruments or larger loads
Tabletop “Class B” autoclaves with a bigger chamber which are 29 Liters 45 Liters, can provide the sterilization solution for bigger size instruments or bigger loads.
Overview
- Benchtop steam sterilizers with Class B pre & post vacuum type, complies with the European standard EN13060
- They have bigger chamber, can provide the sterilization solution for bigger size instruments or bigger loads, such as the instrument kit for dental implants etc.
Excellent Performance
- Full automatic table top autoclave with pre-set programs
- Electronic door locking
- Complies with European standard EN13060
- In built independent rapid steam generator ensures quick sterilization cycles
- Rapid post sterilization vacuum drying
- With one flash water tank and used one water tank
- LCD screen displaying time, temperature and pressure, process alert and instrument conditions
- User friendly interface
- Self-initialization when work in different altitude
- Choice of language setting including: English, Spanish, German, Polish, French, Hungarian, Romanian, Dutch, Lithuanian, Lavian, Czech, Italian and Chinese
Safety and Monitoring
- Double safety protection system prevents the cycle from starting if the door is not correctly locked
- This system also prevents the door from opening if the pressure inside the chamber is not equal to atmospheric pressure outside the chamber
- Pressure safety valve prevents over pressure in the chamber and steam generator
- The power is automatically cut if a short circuit or electrical fault occurs
- Able to detect and identify the exact cause of any problem, and to issue a specific error code to the operator
- Water level main switch in the main tank controlling the min and max water level
- Automatic warning for maintenance
Documentation
- Printer (option): an internal printer is optional for all ICANCLAVE autoclaves
- USB port (option): it is used for connecting a USB stick, all sterilization data are automatically written in the USB stick, and can be ready directly in any PC and stored electronically
- Internal Memory: the last 20 cycles can be automatically stored in the autoclave system, which can be printed out at any time
Technical specifications
Order Code |
IC/T/000229 |
IC/T/000245 |
Chamber Dimension |
Ø 319x417 mm |
Ø319x617 mm |
Chamber Volume |
29 L |
45 L |
Number of Trays |
3 |
3 |
Material of Trays |
Stainless steel |
Stainless steel |
Size of Trays |
240x340 mm |
240x540 mm |
Voltage, Freq. |
220/110V, 50/60Hz |
220/110V, 50/60Hz |
POWER |
2300 w |
3800 w |
Overall Dimension (WxHxD) |
640x640x560 mm |
840x640x560 mm |
Weight |
95 kgs |
115 kgs |
Cycles
Program |
Temp. [·C] |
Pressure [Mpa] |
Sterilization time [min.] |
SOLID |
134 |
210 |
4 |
|
121 |
110 |
20 |
LIQUID |
134 |
210 |
10 |
|
121 |
110 |
30 |
WRAPPED |
134 |
210 |
10 |
|
121 |
110 |
30 |
TEXTILE |
134 |
210 |
10 |
|
121 |
110 |
30 |
PRION |
134 |
210 |
18 |
DRYNG |
- |
- |
- |
B&D TEST |
134 |
210 |
3.5 |
HELIX TEST |
134 |
210 |
3.5 |
VACUUM TEST |
- |
- |
- |
Standards
- 93/42/EEC Directive on Medical Devices
- 97/23/EEC Directive on Pressure Equipment
- EN 13060 Small Steam Sterilizers
- EN 61326, EN 61000 Electromagnetic Compatibility
- EN 61010-1, UL 61010-1Safety requirements
- EN 61010-2-040, IEC 61010-2-040. Specific requirements for steam sterilizers used for treatment of medical materials