NEYA 16 R Refrigerated 16.000 rpm
Benchtop refrigerated centrifuge maximum capacity 4×175 ml, maximum speed 16.000 rpm, 10 Programs
Refrigerated centrifuge "High-Speed" mod. NEYA 16 R, 4x175 ml, 16.000 rpm, 10 programs.

Features
Automatic rotor detection, check of presence of accessories and compatibility with maximum speed.
- Safety speed limiter function
- 10 storable programs with protection function
- Setting of speed in RPM and RCF
- Short Spin function and precool (NEYA 16R)
- Temperature range from -10°C to +40°C (NEYA 16R)
- Controlled by microprocessor and backlit color LCD display with contemporary visualization of all parameters
- Digital adjustment of acceleration and deceleration levels
- Stainless steel internal bowl with optimal height for load and download of samples
- Imbalance detection system with automatic functioning stop to avoid accidents
Automatic locking system of the lid, safety opening of the lid in case of absence of electric power
- Brushless motor maintenance free and no deposits
Technical data
Model |
NEYA 16 R Refrigerated "High-Speed" |
Code |
RM/4010/0342 |
Maximum capacity |
4×175 ml (swing out) - 6×100 ml (fixed angle) |
Maximum speed |
4.500 rpm (swing out) - 16.000 rpm (fixed angle) |
Setting RPM |
YES |
Setting RCF |
YES |
Display RCF |
YES |
Timer |
00:30¸99:50 (mm:ss) / continuos mode |
Date and time |
YES |
Acceleration levels |
0÷9 (0= min - 9=max) |
Deceleration levels |
0÷9 (0= min - 9=max) |
Temperature range |
-10 ÷ +40°C / +14 ÷ +114°F |
Pre-cool function |
YES |
Display temperature |
YES (°C and °F) |
Spin function |
YES |
Programs |
10 programs with protection function |
Indication of rotor |
YES |
Noise |
55 dB |
Directive / Standard |
IEC 1010-1; IEC 1010-2-020 |
External dimensions |
730×640×330 mm |
Weight |
70 kg |
Voltage |
220 +/- 10% V / 50 Hz |
Power |
750 W |
Rotors
This table will show just some rotor configurations; to explore all the rotors, accessories, inserts, visit the page of RM Series rotors.
Codice |
Descrizione |
RM/4010/0502 |
Swing rotor SB 4-175 with 4 bucket and BIO SAFE caps; Kit composed by swing out rotor S 4-175, 4 buckets B 175 and 4 Biosafe lids L 175. Maximum capacity 4x175 ml; Maximum speed 4.500 rpm; Maximum acceleration 3.600 xg |
RM/4010/0522 |
Swing rotor S 6-96 MP for MICROPLATE with 2 supports. Swing out rotor for microplates. Max plates for tray 3. Max plates for rotor 6. Max speed: 3.200 rpm. Max acceleration: 1.950 xg |
RM/4010/0802 |
Fixed angle rotor A 32-15 ( 32 tubes 15 ml ). Fixed angle rotor 37°. N° tubes for rotor 32. Max speed: 4.500 rpm. Max acceleration: 3.280 xg. Including n 32 metal bucket B 15F with conical Falcon tube type 15 ml. |
RM/4010/0812 |
Fixed angle rotor A 8-50 ( 8 tubes 50 ml ). Fixed angle rotor 37°. N° tubes for rotor 8. Max speed: 6.000 rpm. Max acceleration: 4.800 xg. Including n 8 metal bucket B 50XF with Falcon tube type conical 50ml. |
RM/4010/0822 |
Fixed angle rotor A 6-100 ( 6 tubes 100 ml ). Fixed angle rotor 37°. N° tubes for rotor 6. Max speed: 5.000 rpm. Max acceleration: 3.330 xg. Including n 6 metal bucket B 100-P for rotor A 6-100 with plastic tube round bottom 100 ml. |
RM/4010/0862 |
Fixed angle rotor A 6-50 HIGH SPEED (6 tubes 50 ml ). lid included |
RM/4010/0832 |
Fixed angle rotor A 12-5 (12 microtubes 5 ml Eppendorf).lid included |
RM/4010/0842 |
Fixed angle rotor A 24-2 (24 microtubes 1,5-2 ml ).lid included. |
RM/4010/0852 |
Fixed angle rotor A 36-05 (36 microtubes 0,5 ml ).lid included. |
RM/4010/0872 |
Fixed angle rotor PCR 4-8 (4 strip with 8 cuvets each for PCR ).lid included. |
RM/4010/0882 |
Rotor for Hematocrit HE 24-75 (24 capillaries).lid included. |