Portable multi-parameter Mod. 907

One device allows connecting Memosens pH, conductivity or oxygen sensors (also optical oxygen!).

Multi-parameter portable device for Memosens pH, conductivity and oxygen sensors.

The first portables with Memosens technology for measuring pH, conductivity and oxygen. Also for hazardous areas.

Multiparameter. One device allows connecting Memosens pH, conductivity or oxygen sensors (also optical oxygen!).

Intuitive. The 907 Multi features an intuitive user interface with a plain text menu in 6 selectable languages. The high-resolution transflective color graphic display is suitable for use in sunlight and with its scratch-proof mineral glass can still be read perfectly after many years.

Communicative. Data logger entries can be shown on the display using an extensive recorder function. The standard micro USB interface enables

  • connection to a PC
  • software updates
  • charging the Li-ion battery

The data logger records up to 10,000 values.
The included Paraly SW 112 software enables easy management of recorded values.

Many new features distinguish the 907 Multi for use in the pharmaceutical and biotech field. These include

  • a new pH calibration procedure with a fixed process cycle
  • multi-level user management with access control
  • direct assignment of Memosens sensors to the device, which increases safety in operation

The 907 Multi pH Multiparameter portable meter with additional analog pH measurement capability.

For all digital Memosens pH, conductivity and oxygen sensors and for analog pH electrodes.

The only portable device for all Memosens parameters. Also for conventional analog sensors.

The powerful Li-ion battery can be charged in the device via USB. The clear network diagram provides an at-a-glance view of the sensor condition.

Comprehensive data logger The following logging types can be selected:

  • Manual logging
  • Time-controlled logging at a fixed interval
  • Signal-controlled logging of measured value and temperature – Combined time- and signal-controlled logging
  • Threshold-controlled logging with pre-trigger

The data logger for up to 10,000 entries records point of measurement, annotation, sensor ID, sensor serial number (Memosens), primary value, temperature, time stamp and device status.

User-friendly software

Mod.  907 proves that a high level of functionality and easy use are not mutually exclusive. It proceeds step by step through the calibration procedure.

Technical terms are clearly explained in the contextsensitive help.

Facts

  • High-resolution color graphic display
  • Transflective and sunlight readable
  • Li-ion battery
  • Micro USB port and Paraly SW 112 software
  • A sensor quiver protects the sensor from damage and drying out
  • The high-performance polymer housing ensures low water consumption and high impact resistance
  • Intelligent data logger with 10,000 entries and graphical representation
  • Memosens sensors and analog pH sensors can be used on one device (e.g flat-membrane sensors)
  • IP 67 / IP 66 protection
  • The mineral glass display is perfectly readable even after years

Specifications

pH/mV input (analog) pH socket, DIN 19 262 (13/4 mm)
pH range -2 … 16
Decimal places*) 2 or 3
Input resistance 1 x 1012 Ω (0 … 35 °C)
Input current 1 x 10-12 A (at RT, doubles every 10 K)
Measuring cycle Approx. 1 s
Measurement error1,2,3) < 0.01 pH, TC < 0.001 pH/K
mV range -1300 … +1300 mV
Measuring cycle Approx. 1 s
Measurement error1,2,3) < 0.1 % meas.val. + 0.3 mV TC < 0.03 mV/K
Temperature input 2 x 4 mm dia. for integrated or separate temperature detector
Measuring ranges NTC 30 kΩ -20 … +120 °C Pt 1000 -40 … +250 °C
Measuring cycle Approx. 1 s
Measurement error1,2,3) < 0.2 K (Tamb = 23 °C); TC < 25 ppm/K
Memosens pH input (also ISFET) M8 socket, 4 pins, for Memosens lab cable
Display ranges4)
pH -2.000 … +16.000
mV -2000 … +2000 mV
Temperature -50 … +250 °C
Memosens ORP input M8 socket, 4 pins, for Memosens lab cable
Display ranges4)
mV -2000 … +2000 mV
Temperature -50 … +250 °C
Sensor standardization*) ORP calibration (zero adjustment)
Permissible calibration range ΔmV (offset) -700 … +700 mV
Sensor standardization*) pH calibration
Operating modes*) Calimatic Calibration with automatic buffer recognition
Manual Manual calibration with entry of individual buffer values
Data entry Data entry of zero and slope
Permissible calibration range Zero point 6 ... 8 pH With
ISFET: -750 … +750 mV Operating point (asymmetry)
Slope Approx. 74 … 104 %
Calibration timer* Interval 1 … 99 days, can be switched off
Sensoface Provides information on the sensor condition Evaluation of zero/slope, response, calibration interval
Conductivity input, Memosens M8 socket, 4 pins, for Memosens lab cable or
Measuring range SE 215 MS sensor 10 µS/cm … 20 mS/cm
Measuring cycle Approx. 1 s
Temperature compensation Linear 0 … 20 %/K, reference temp. adjustable nLF: 0 … 120 °C NaCl HCl (ultrapure water with traces) NH3 (ultrapure water with traces) NaOH (ultrapure water with traces)
Display resolution5) (autoranging) Conductivity
0.001 μS/cm (c < 0.05 cm–1)
0.01 μS/cm (c = 0.05 … 0.2 cm–1)
0.1 μS/cm (c > 0.2 cm–1)
Resistivity 00.00 … 99.99 MΩ • cm
Salinity 0.0 … 45.0 g/kg (0 … 30 °C)
TDS 0 … 1999 mg/l (10 … 40 °C)
Concentration 0.00 … 9.99 % by wt
Concentration determination NaCl 0.00 ... 9.99 % by wt (0 ... 60 °C)
HCl 0.00 ... 9.99 % by wt (-20 ... 50 °C)
NaOH 0.00 ... 9.99 % by wt (0 ... 100 °C)
H2SO4 0.00 ... 9.99 % by wt (-17 ... 110 °C)
HNO3 0.00 ... 9.99 % by wt (-17 ... 50 °C)
Sensor standardization Cell constant Input of cell constant with simultaneous display of conductivity value and temperature
Input of solution Input of conductivity of the calibration solution with simultaneous display of cell constant and temperature
Auto Automatic determination of the cell constant with KCl solution or NaCl solution
Memosens input, oxygen M8 socket, 4 pins, for Memosens lab cable
Display ranges4)
Saturation 0.000 ...1000.0 %
Concentration 000 μg/l … 100.00 mg/l
Partial pressure 0.0 … 2000 mbars
Temperature meas. range4) -20 … 150 °C
Sensor standardization Automatic calibration in air, humidity adjustable
Zero calibration
Storage in quiver
Connections 2x socket, 4 mm dia., for separate temp. detector
1x M8 socket, 4 pins, for Memosens lab cable
1x micro USB-B for data transmission to PC
1x pH socket, to DIN 19262
User interface Straightforward menu navigation with graphic icons and detailed operating instructions in plain text
Languages German, English, French, Spanish, Italian, Portuguese, Russian
Data logger 10,000 memory locations
Recording Manual, interval- and/or event-controlled with limit value and pre-trigger, management of tag numbers and annotations
MemoLog calibration data logger  (Memosens only) Up to 100 Memosens calibration records can be saved
– Recording viewable on the display
– directly retrievable via MemoSuite (USB)
Manufacturer, sensor type, serial no., zero, slope, calibration date
Communication USB 2.0
Profile HID, driverless installation
Usage Data exchange and configuration via Paraly SW 112 software
Diagnostics functions Sensor data (only Memosens) Manufacturer, sensor type, serial number, wear, operating time
Calibration data Calibration date, zero, slope
Device self-test Automatic memory test (FLASH, EEPROM, RAM)
Device data Device type, software version, hardware version
Data retention Parameters, calibration data > 10 years
EMC EN 61326-1 (General Requirements) Emitted interference Class B (residential area) Immunity to interference Industry EN 61326-2-3 (Particular Requirements for Transmitters)
RoHS conformity According to directive 2011/65/EU
Power supply 4x AA batteries
4x rechargeable NiMH batteries
1x Li-ion battery, USB chargeable
Nominal operating conditions Ambient temperature -10 … +55 °C
Transport/Storage temp. -25 … +70 °C
Relative humidity 0 … 95 %, short-term condensing allowed
Housing Material PA12 GF30 + TPE
Ingress protection IP66/67 with pressure compensation
Dimensions Approx. (132 x 156 x 30) mm
Weight Approx. 500 g

*) user-defined
1) According to EN 60746-1, at nominal operating conditions
2) ± 1 count
3) Plus sensor error
4) Ranges depending on Memosens sensor
5) c = cell constant