pH/mV input (analog) |
pH socket, DIN 19 262 (13/4 mm) pH range -2 … 16 Decimal places*) 2 or 3 Input resistance 1 x 1012 Ω (0 … 35 °C) Input current 1 x 10-12 A (at RT, doubles every 10 K) Measuring cycle Approx. 1 s Measurement error1,2,3) < 0.01 pH, TC < 0.001 pH/K mV range -1300 … +1300 mV Measuring cycle Approx. 1 s Measurement error1,2,3) < 0.1 % meas.val. + 0.3 mV TC < 0.03 mV/K |
Temperature input |
2 x 4 mm dia. for integrated or separate temperature detector Measuring ranges NTC 30 kΩ -20 … +120 °C Pt 1000 -40 … +250 °C Measuring cycle Approx. 1 s Measurement error1,2,3) < 0.2 K (Tamb = 23 °C); TC < 25 ppm/K |
Memosens pH input (also ISFET) |
M8 socket, 4 pins, for Memosens lab cable Display ranges4) pH -2.000 … +16.000 mV -2000 … +2000 mV Temperature -50 … +250 °C |
Memosens ORP input |
M8 socket, 4 pins, for Memosens lab cable Display ranges4) mV -2000 … +2000 mV Temperature -50 … +250 °C Sensor standardization*) ORP calibration (zero adjustment) Permissible calibration range ΔmV (offset) -700 … +700 mV |
Sensor standardization*) |
pH calibration |
Operating modes*) |
Calimatic Calibration with automatic buffer recognition Manual Manual calibration with entry of individual buffer values Data entry Data entry of zero and slope |
Permissible calibration range |
Zero point 6 ... 8 pH With ISFET: -750 … +750 mV Operating point (asymmetry) Slope Approx. 74 … 104 % |
Calibration timer* |
Interval 1 … 99 days, can be switched off |
Sensoface |
Provides information on the sensor condition Evaluation of zero/slope, response, calibration interval |
Conductivity input, Memosens |
M8 socket, 4 pins, for Memosens lab cable or Measuring range SE 215 MS sensor 10 µS/cm … 20 mS/cm Measuring cycle Approx. 1 s Temperature compensation Linear 0 … 20 %/K, reference temp. adjustable nLF: 0 … 120 °C NaCl HCl (ultrapure water with traces) NH3 (ultrapure water with traces) NaOH (ultrapure water with traces) |
Display resolution5) (autoranging) |
Conductivity 0.001 μS/cm (c < 0.05 cm–1) 0.01 μS/cm (c = 0.05 … 0.2 cm–1) 0.1 μS/cm (c > 0.2 cm–1) Resistivity 00.00 … 99.99 MΩ • cm Salinity 0.0 … 45.0 g/kg (0 … 30 °C) TDS 0 … 1999 mg/l (10 … 40 °C) Concentration 0.00 … 9.99 % by wt |
Concentration determination |
NaCl 0.00 ... 9.99 % by wt (0 ... 60 °C) HCl 0.00 ... 9.99 % by wt (-20 ... 50 °C) NaOH 0.00 ... 9.99 % by wt (0 ... 100 °C) H2SO4 0.00 ... 9.99 % by wt (-17 ... 110 °C) HNO3 0.00 ... 9.99 % by wt (-17 ... 50 °C) |
Sensor standardization |
Cell constant Input of cell constant with simultaneous display of conductivity value and temperature Input of solution Input of conductivity of the calibration solution with simultaneous display of cell constant and temperature Auto Automatic determination of the cell constant with KCl solution or NaCl solution |
Memosens input, oxygen |
M8 socket, 4 pins, for Memosens lab cable Display ranges4) Saturation 0.000 ...1000.0 % Concentration 000 μg/l … 100.00 mg/l Partial pressure 0.0 … 2000 mbars Temperature meas. range4) -20 … 150 °C |
Sensor standardization |
Automatic calibration in air, humidity adjustable Zero calibration |
Storage |
in quiver |
Connections |
2x socket, 4 mm dia., for separate temp. detector 1x M8 socket, 4 pins, for Memosens lab cable 1x micro USB-B for data transmission to PC 1x pH socket, to DIN 19262 |
User interface |
Straightforward menu navigation with graphic icons and detailed operating instructions in plain text |
Languages |
German, English, French, Spanish, Italian, Portuguese, Russian |
Data logger |
10,000 memory locations Recording Manual, interval- and/or event-controlled with limit value and pre-trigger, management of tag numbers and annotations |
MemoLog calibration data logger (Memosens only) |
Up to 100 Memosens calibration records can be saved – Recording viewable on the display – directly retrievable via MemoSuite (USB) Manufacturer, sensor type, serial no., zero, slope, calibration date |
Communication |
USB 2.0 Profile HID, driverless installation Usage Data exchange and configuration via Paraly SW 112 software |
Diagnostics functions |
Sensor data (only Memosens) Manufacturer, sensor type, serial number, wear, operating time Calibration data Calibration date, zero, slope Device self-test Automatic memory test (FLASH, EEPROM, RAM) Device data Device type, software version, hardware version |
Data retention |
Parameters, calibration data > 10 years |
EMC |
EN 61326-1 (General Requirements) Emitted interference Class B (residential area) Immunity to interference Industry EN 61326-2-3 (Particular Requirements for Transmitters) |
RoHS conformity |
According to directive 2011/65/EU |
Power supply |
4x AA batteries 4x rechargeable NiMH batteries 1x Li-ion battery, USB chargeable |
Nominal operating conditions |
Ambient temperature -10 … +55 °C Transport/Storage temp. -25 … +70 °C Relative humidity 0 … 95 %, short-term condensing allowed |
Housing |
Material PA12 GF30 + TPE Ingress protection IP66/67 with pressure compensation Dimensions Approx. (132 x 156 x 30) mm Weight Approx. 500 g |